Effect of both pH and Organic Content on a Separation of Sugars, Amino Acids, and Carboxylic Acids


In mixed-mode HILIC chromatography, selectivity of separation can be adjusted by amount of acetonitrile, amount of buffer and buffer pH. Buffer concentration and pH will affect retention of ionizable compounds to a different degree. Retention of neutral compounds can be adjusted by the amount of acetonitrile. Carboxylic acid, three amino acids and two sugars are separated by combination of HILIC and ion-exchange mechanisms. Compounds can be monitored by ELSD, Corona (CAD), LC/MS or low UV. UV-transparent mobile phase /buffer is required for UV monitoring of this mixed-mode separation. This HPLC method can be adopted as general approach for analysis of sugars, amino acids and carboxylic acids.

Condition

Column Obelisc N, 4.6×150 mm, 5 µm, 100A
Mobile Phase MeCN/H2O
Buffer AmAc
Flow Rate 1.0 ml/min
Detection ELSD

 

Description

Class of Compounds
Drug, Acid, Hydrophilic, Ionizable, Vitamin, Supplements, Amino acid
Analyzing Compounds Succinic Acid, Phenylalanine, Sucrose, Glycine, Aspartic Acid, Raffinose

Application Column

Obelisc N

SIELC has developed the Obelisc™ columns, which are mixed-mode and utilize Liquid Separation Cell technology (LiSC™). These cost-effective columns are the first of their kind to be commercially available and can replace multiple HPLC columns, including reversed-phase (RP), AQ-type reversed-phase, polar-embedded group RP columns, normal-phase, cation-exchange, anion-exchange, ion-exclusion, and HILIC (Hydrophilic Interaction Liquid Chromatography) columns. By controlling just three orthogonal method parameters - buffer concentration, buffer pH, and organic modifier concentration - users can adjust the column properties with pinpoint precision to separate complex mixtures.

Select options
Application Analytes:
Aspartic Acid
Phenylalanine
Succinic Acid

Application Detection:
ELSD Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.