Amino Acids Analysis in Acid Gradient Condition |
 | Application Notes:
11 underivatized amino acids (aspartic acid, glutamic acid, alanine, valine, methionine, isoleucine, cysteine, phenylalanine, histidine, lysine, and arginine) are separated by a Primesep 100 HPLC column by reversed-phase and ion-exchange mechanisms with LC/MS compatible conditions without the use of ion-pair reagents. The HPLC separation uses a TFA (trifluoroacetic acid) gradient in a mobile phase of water acetonitrile (MeCN, ACN with evaporative light scattering detection (ELSD).
Application Columns:
- Primesep® 100 is a mixed-mode HPLC column with general hydrophobic and ion-exchange properties. The Primesep 100 stationary phase combines a hydrophobic chain and an acidic functional group as a cation exchanger. Acids, bases, zwitterions and neutrals can be retained and separated by Primesep 100 columns. Primesep 100 retains hydrophobic compounds by reversed-phase retention alone or a combination of reversed-phase and ion-exchange retention. Primesep 100 separates underivatized amino acids and amines by combining reversed-phase and ion-exchange mechanisms. In this case the acidic functional group acts as an embedded ion-pairing reagent. Organic acids are separated by reversed phase and ion exclusion. Inorganic cations can be separated by cation exchange. Methods can be developed on Primesep columns that are fully compatible with detectors such as UV, RI, conductivity, fluorescence, evaporative light scattering (ELSD) as well as mass spectrometers (LC/MS) and preparative chromatography (prep HPLC) systems.
Application Analytes:
Application Detection:
|
Hydrophobic and Hydrophilic Compound Separation |
 | Application Notes:
Primesep 100 separates a mixture of polar and nonpolar compounds in one analytical run. The amino acid cysteine; amino acid derivatives L-cystine, 2,2-dimethylcystine, and 2-methylcysteine; the polar acid benzoic acid; and the nonpolar neutral toluene are separated by a gradient using a combination of polar and hydrophobic interactions. The separation method uses a mobile phase mixture of water, acetonitrile (MeCN, ACN) and sulfuric acid (H2SO4) with UV detection at 210 nm.
Application Columns:
- Primesep® 100 is a mixed-mode HPLC column with general hydrophobic and ion-exchange properties. The Primesep 100 stationary phase combines a hydrophobic chain and an acidic functional group as a cation exchanger. Acids, bases, zwitterions and neutrals can be retained and separated by Primesep 100 columns. Primesep 100 retains hydrophobic compounds by reversed-phase retention alone or a combination of reversed-phase and ion-exchange retention. Primesep 100 separates underivatized amino acids and amines by combining reversed-phase and ion-exchange mechanisms. In this case the acidic functional group acts as an embedded ion-pairing reagent. Organic acids are separated by reversed phase and ion exclusion. Inorganic cations can be separated by cation exchange. Methods can be developed on Primesep columns that are fully compatible with detectors such as UV, RI, conductivity, fluorescence, evaporative light scattering (ELSD) as well as mass spectrometers (LC/MS) and preparative chromatography (prep HPLC) systems.
Application Analytes:
Application Detection:
|
Complex Mixture of Acids, Bases, Amino Acids, and Neutral Compounds |
 | Application Notes:
Primesep 100 separates a mixture of amino acids (tyrosine, phenylalanine), organic acids (benzoic acid, mandelic acid), amines (benzylamine, pyridine), and neutrals (benzonitrile, toluene) in one HPLC run by combining reversed-phase, cation-exchange, and polar interactions. The method is tunable and peak order can be changed significantly by adjusting acetonitrile and trifluoroacetic acid concentrations. The separation method uses a mobile phase mixture of water, acetonitrile (MeCN, ACN) and trifluoracetic acid (TFA) and compatible with UV, mass spec (LC/MS) and evaporative light scattering (ELSD) detection.
Application Columns:
- Primesep® 100 is a mixed-mode HPLC column with general hydrophobic and ion-exchange properties. The Primesep 100 stationary phase combines a hydrophobic chain and an acidic functional group as a cation exchanger. Acids, bases, zwitterions and neutrals can be retained and separated by Primesep 100 columns. Primesep 100 retains hydrophobic compounds by reversed-phase retention alone or a combination of reversed-phase and ion-exchange retention. Primesep 100 separates underivatized amino acids and amines by combining reversed-phase and ion-exchange mechanisms. In this case the acidic functional group acts as an embedded ion-pairing reagent. Organic acids are separated by reversed phase and ion exclusion. Inorganic cations can be separated by cation exchange. Methods can be developed on Primesep columns that are fully compatible with detectors such as UV, RI, conductivity, fluorescence, evaporative light scattering (ELSD) as well as mass spectrometers (LC/MS) and preparative chromatography (prep HPLC) systems.
Application Analytes:
Application Detection:
|
Bufferless Ion Separation (BLIS™) Chromatography of Amino Acids (2) |
 | Application Notes:
Application Columns:
- Primesep® 200 is a mixed-mode HPLC column with general hydrophobic and ion-exchange properties. The Primesep 200 stationary phase combines a hydrophobic chain and a weakly acidic functional group as a cation exchanger. Acids, bases, zwitterions and neutrals can be retained and separated by Primesep 200 columns. Primesep 200 retains hydrophobic compounds by reversed-phase retention alone or a combination of reversed-phase and ion-exchange retention. Primesep 200 separates underivatized amino acids, amines, isomers and structurally related compounds by combining reversed phase and ion exchange. In this case the acidic functional group acts as an embedded ion-pairing reagent. Organic acids and diacids are separated by reversed phase and ion exclusion. Methods can be developed on Primesep columns that are fully compatible with detectors such as UV, RI, conductivity, fluorescence, evaporative light scattering (ELSD) as well as mass spectrometers (LC/MS) and preparative chromatography (prep HPLC) systems.
Application Analytes:
Application Detection:
| |
|