Primesep 200 separates the very hydrophilic polyamines, 1,3-diaminopropane, N-(3-aminopropyl)-1,3-diaminepropane, N,N-bis-(3-aminopropyl)-1,3-diaminepropane. The retention and resolution are affected by either a TFA concentration gradient, or a pH gradient from pH 4 to 3. This flexibility is obtained by simple mobile phases of water, acetonitrile (MeCN, ACN), ammonium formate, ammonium acetate or trifluoracetic acid (TFA) with evaporative light scattering detection (ELSD).
Application Columns:
- Primesep® 200 is a mixed-mode HPLC column with general hydrophobic and ion-exchange properties. The Primesep 200 stationary phase combines a hydrophobic chain and a weakly acidic functional group as a cation exchanger. Acids, bases, zwitterions and neutrals can be retained and separated by Primesep 200 columns. Primesep 200 retains hydrophobic compounds by reversed-phase retention alone or a combination of reversed-phase and ion-exchange retention. Primesep 200 separates underivatized amino acids, amines, isomers and structurally related compounds by combining reversed phase and ion exchange. In this case the acidic functional group acts as an embedded ion-pairing reagent. Organic acids and diacids are separated by reversed phase and ion exclusion. Methods can be developed on Primesep columns that are fully compatible with detectors such as UV, RI, conductivity, fluorescence, evaporative light scattering (ELSD) as well as mass spectrometers (LC/MS) and preparative chromatography (prep HPLC) systems.
Application Analytes:
Application Detection:
| |
|