Separation and Retention of Vitamins in Mixed-Mode HPLC |
 | Application Notes:
Primesep B and B2 columns separate Vitamins C (ascorbic acid) and B2 (riboflavin) with tunable selectivity. Both columns allow peak order reversal by simply changing the acetonitrile concentration in the mobile phase. Vitamin B2 is especially sensitive to acetonitrile concentration. The HPLC separation uses a mobile phase of water, acetonitrile (MeCN, ACN) and either trifluoroacetic acid (TFA) or acetic acid (HOAc) with UV detection at 250 nm.
Application Analytes:
Application Detection:
|
Separation of Two Vitamins: Different Polarity - Isocratic Methods |
 | Application Notes:
Primesep B and B2 columns separate Vitamins C (ascorbic acid) and B2 (riboflavin) with tunable selectivity. Peak order reversal is exhibited on these two columns with the same mobile phases due to their different polarity. The HPLC separation uses a mobile phase of water, acetonitrile (MeCN, ACN) and either formic or acetic acid (HOAc) with UV detection at 250 nm.
Application Analytes:
Application Detection:
|
HPLC Separation of Vitamin C and Vitamin B2 |
 | Application Notes:
Vitamin C (ascorbic acid) and Vitamin B2 (riboflavin) are water-soluble vitamins that are not stored in the body and must be replaced each day. Primesep D easily resolves Vitamins C and B2 and the elution order can be reversed by adjusting acetonitrile and TFA concentrations in the mobile phase. The HPLC separation uses a mobile phase of water, acetonitrile (MeCN, ACN) and trifluoroacetic acid (TFA) and UV detection at 250 nm.
Application Analytes:
Application Detection:
|
Separation of Vitamins in Mixed-Mode HPLC |
 | Application Notes:
Application Analytes:
Application Detection:
|
HILIC Separation of Common Preservatives - Citric Acid, Ascorbic Acid and EDTA |
 | Application Notes:
Citric acid, ascorbic acid and EDTA are commonly used in food and pharmaceutical industry as preservatives. These compounds are very polar in nature. They are weak organic acids with limited UV activity. Retention and separation is achieved on HILIC mixed-mode Obelisc N column. All three compounds are retained by combination of strong HILIC and strong anion-exchange mechanisms. Separation can be monitored by ELSD, LC/MS, UV or Corona CAD. In contrast to other HILIC column, Obelisc N has two ionizable groups basic and acidic which provide ion-exchange interaction in addition to hydrophilic interaction. This allows to use less acetonitrile for HILIC separation.
Application Analytes:
Application Detection:
| |
|