HPLC Determination of Carbocisteine on Primesep 100 Column

Carbocisteine Carbocisteine, or carbocysteine, is a drug designed to clear the mucus from the upper and lower airways. It’s primarily available in tablet form. It can be retained on a Primesep 100 Column in about 5 minutes. Primesep 100 is a reverse-phase (RP) mixed-mode column with embedded acidic ion-pairing groups.

Condition

Column Primesep 100, 4.6x150 mm, 5 µm, 100A
Mobile Phase MeCN/H2O - 10/90%
Buffer H2SO4 - 0.05 %
Flow Rate 1.0 ml/min
Detection UV, 200 nm
 

Description

Class of Compounds Acid, Hydrophilic, Ionizable, Carboxylic acid, Drug, Carbocyclic, Amino acid
Analyzing Compounds Carbocisteine
 


Application Analytes:

Carbocisteine
S-(Carboxymethyl)-L-cysteine