HPLC Determination of Rifampicin on Newcrom A column

HPLC.cloud View on hplc.cloud Rifampicin is an antibiotic used alongside other antibiotics to treat several types of bacterial infections like tuberculosis and Legionnaires’ disease. It works by inhibiting the bacteria’s RNA production and is on the World Health Organisation’s (WHO) List of essential medicines. It can be retained in HPLC on Newcrom A mixed-mode column. The retention characteristics can be controlled by the organic modifier used in the mobile phase, with methanol (MeOH) having a longer retention time in comparison to acetonitrile (ACN). UV detection at 275nm.

Condition

Column Newcrom A, 4.6x50 mm, 3 µm, 100A
Mobile Phase MeOH, MeCN
Buffer H2SO4 - 0.5%
Flow Rate 1.0 ml/min
Detection UV 275 nm
 

Description

Class of Compounds Hydrophobic, Drugs, Antibiotics
Analyzing Compounds Rifampicin


Application Analytes:

Rifampicin