Applications:
HPLC Method For Analysis Of Trimetazidine Hydrochloride on Primesep 100 Column
March 22, 2022
HPLC Method for Trimetazidine on Primesep 100 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Trimetazidine.
Trimetazidine (TMZ) is a medication with the chemical formula C14H22N2O3. It is taken to treat angina and other heart conditions. It inhibits the mitochondrial enzyme long-chain-3-ketoacyl-CoA thiolase which decreases fatty acids and shifts substrate utilization towards glucose oxidation. There is a debate within the sports and scientific communities about whether or not TMZ is a performance-enhancing drug. The World Anti-Doping Agency considers it to be a hormone and metabolic modulator.
Trimetazidine can be retained and analyzed using the Primesep 100 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a sulfuric acid buffer. Detection is performed using UV.
| Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | MeCN/H2O – 60/40% |
| Buffer | H2SO4 – 0.4% |
| Flow Rate | 1.0 ml/min |
| Detection | UV, 200 nm, |
| Class of Compounds |
Drug |
| Analyzing Compounds | Trimetazidine |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended
Separation of Trimetazidine on Newcrom R1 HPLC column
February 16, 2018
Trimetazidine can be analyzed by this reverse phase (RP) HPLC method with simple conditions. The mobile phase contains an acetonitrile (MeCN), water, and phosphoric acid. For Mass-Spec (MS) compatible applications the phosphoric acid needs to be replaced with formic acid. Smaller 3 µm particles columns available for fast UPLC applications. This liquid chromatography method is scalable and can be used for isolation impurities in preparative separation. It also suitable for pharmacokinetics.
Application Column
Newcrom R1
The Newcrom columns are a family of reverse-phase-based columns. Newcrom A, AH, B, and BH are all mixed-mode columns with either positive or negative ion-pairing groups attached to either short (25 Å) or long (100 Å) ligand chains. Newcrom R1 is a special reverse-phase column with low silanol activity.