Newcrom R1

SIELC introduces a new single ended column design

Newcrom - single ended column design

 

View brochure
Newcrom brochureNewcrom brochure

 

 

 

 

 

 

 

For as long as people have done HPLC they have used the same column format, the same method of attaching the column to the HPLC, and they have  experienced the same problems:

  • Clogging at the tube-column junction
  • Leakage around the high pressure solvent connections
  • Installation problems with tubing to achieve zero dead volume
  • Selecting the proper tube length and ID
  • Sorting through many available high pressure fittings

To address these issues  a new column design was developed. This column and corresponding adapter entirely eliminate the need

for any high-pressure fittings or tubing as well as minimizing all possible dead volumes. Furthermore, if a leak ever occurs in the high-pressure column inlet, the mobile phase is contained within the column adapter (no external leakage).
You need to install a column adapter once on your HPLC. Then without any tools you can install a column which never leaks, it is easy to replace and can be changed in seconds.  The column hardware was  designed specifically to be used in any HPLC instrument. It can be installed in most column heaters as well.

Newcrom R1 columns are available in all standard dimensions

Column IDs: 22 mm, 10 mm, 4.6 mm, 3.2 mm, 2.1 mm
Column Lengths: 250, 150, 100, 50, 25, 10 mm
Particles: 5 µm and 3 µm
Pores: 100 A

This column is available in two hardware versions traditional two ended and new SIELC hardware with only one end.

Single ended column adapter

Newcrom R1 stationary phase is generally stable at basic pH values.
Recommended pH is from 1.0 to 10.0. Newcrom C18 pH range
As an example, we perform a test where the column was operated at pH 10.6 injecting a model mixture of a neutral, basic, and acidic compounds. This experiment took over 15 hrs and after almost 200 injections no significant lost or changes of retention were observed.
However despite the fact that this stationary phase is stable at high pH it is recommended for column storage to use acidic mobile phase. A mobile phase composition with 0.1% phosphoric or formic acid in 50/50 mixture of MeCN/H2O should be an ideal long and short term storage solution.

Order Newcrom R1


 

Newcrom R1 Application Examples

HPLC Determination of 6-methyl-5-hepten-2-one

HPLC Separation of Mixture of Mycophenolate mofetil and Mycophenolic Acid
HPLC Method for Analysis of Fluticasone Propionate in Nasal Spray
HPLC Method for Analysis of Pantoprazole Sodium in Tablets
HPLC Method for Analysis of Nitroglycerin
HPLC Method for Analysis of Quetiapine in Tablets
HPLC Method for Analysis of Doxazosin Mesylate in Tablets
HPLC Method for Analysis of Ethyl red
HPLC Method for Analysis of D-alpha-Tocopheryl acetate
HPLC Method For Analysis Of Alprazolam
HPLC Method For Analysis Of Minocycline hydrochloride
HPLC Method for Analysis of Dansyl Amino Acids
HPLC Separation of Acetaminophen, Caffeine and Pyrilamine maleate
HPLC Method for Analysis of Dyes
HPLC Method for Analysis of Metronidazole
HPLC Method for Analysis of Warfarin
HPLC Method for Analysis of Verapamil